Leave inside the sterilization chamber only the tray carrier with the trays. All other
8.
accessories should be positioned in a separate space available to operators.
Never place on the autoclave newspapers, trays, liquid containers, etc.
9.
10. Do not lean on the door when it is open
11. Comply with the minimum overall dimensions shown in PICT.22
(PICT.1 pos..0. – FIG.18) for ensure the minum distance to wall.
12. Connect the overflow connectors to the special drains with the pipes provided
(chapter 5)
13. Always make sure the electrical system to which the autoclave is to be connected is in
conformity with the legislation in force and sized to suit the specifications of the said
appliance.
14. Take the power supply cable supplied and plug the plug on the rear panel of the
autoclave into the female socket (PICT.1 pos..6.)
15. Connect the electric plug to the system and make sure the power supply is 230 Vac -
2000 Va minimum
Do not use extensions, reducers or adapters for connection as this could
NOTA:
cause micro interruptions with consequent generation of alarm signals
16. Turn on the autoclave by pushing the general switch (PICT.2 pos. 17.) and open the
door to the autoclave. Wait a few seconds, there will be two beeps which indicate the
reaching of relative parameters for automatic atmospheric alignment, at the same time
the display will indicate "door open".
Never select a command before hearing the two sound signals, the
NOTA:
autoclave will not accept the chosen programming.
17. Load the pure water tank following the indications in the section entitled USAGE
INSTRUCTIONS (chapter 8).
18. Successively run the test cycle, always following the indications expressed in the
section USAGE INSTRUCTIONS (chapter 8).
19. If the autoclave is installed with the demineralizer system, follow the instructions
contained in the packaging of this device and the demineralizer section
(chapter 14).
20
Use the spacer
.
.