Usage Instruction; Loading Pure Water Tank - Tecno-gaz Europa B evo Mode D'emploi

L'autoclave
Masquer les pouces Voir aussi pour Europa B evo:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

08
After having installed the autoclave, proceed with preparations and use.
8.1
Turn on the autoclave
Press the general switch (PICT.2 pos..17.).
8.2

Loading pure water tank

Connect the water filling pipe (provided) with the connection for water filling pipe and
plastic funnel PICT.14. Then engage the connection
(PICT.2 pos. .13.).
Insert the other end of the pipe with filter in the demineralized or distilled water container.
Press the button
B-PUMP
start.
The pump loads the tank inside the autoclave. If maximum level is not reached within 180
seconds the pump automatically stops, therefore it is necessary to press the button again.
The pump stops automatically when maximum level is reached.
With connection to a deionizer, if the maximum water level has not been reached, the
functioning of the autoclave will be obstructed.
8.3
Manual filling
If you wish to use this method of filling or in the event of malfunctioning of the pump, the
filling of the tank can be carried out manually by the operator in the following way:
Turn on autoclave
Remove the connection cover of (PICT.2 pos..12.)
Insert in this opening the hose connected to the funnel (PICT. 14)
Pour distilled water in the funnel, keeping it higher than the loading top
Add until the led of maximum level turns on (PICT.2 pos. 27.).
Loading completed, twist on the top

USAGE INSTRUCTION

to start the water loading pump and press it until the countdown
ENGLISH
in the
charging hole
21

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Europa b evo 24

Table des Matières