Install the autoclave in an environment suitable for carrying out sterilization
1.
procedures.
The room must be adequately lit and ventilated, as provided for in the directives in
2.
force
Install the autoclave away from heat sources and water splashes.
3.
Position the autoclave on a surface that is able to bear its weight. The minimum
4.
recommended load bearing capacity of the surface is 80 kg.
The supporting surface must be perfectly horizontal and must not have inclinations.
5.
Place the autoclave at a height that allows the user to inspect and easily clean all
6.
parts of the sterilization chamber.
Open the door of the autoclave and remove from inside the sterilization chamber all
7.
packages which contain the single accessories.
INSTALLATION
PICT.22
EUROPA B EVO
590 mm
426 mm (front)
455 mm (rear)
400 mm
EUROPA B EVO 24
610 mm
730 mm
50 mm
410 mm
403 mm (front)
444 mm (rear)
431 mm
ENGLISH
19