Flushing the Cutting Chamber

Cooling Valve
The cooling valve for cooling/cleaning
is located in the back of the cutting
chamber to the left.
Turn the knob
counter-
clockwise to
vertical position
for cooling of
cut-off wheel.
Turn the knob
to regulate the
flow while
flushing the
cutting
chamber.
Turn the knob
clockwise to
vertical position
for flushing the
cutting chamber
with maximum
flow.
18
Unitom-5
Instruction Manual
To ensure a longer lifetime of your Unitom-5, Struers strongly
recommends that you clean the cutting chamber with the cleaning
hose every day.
Point the cleaning hose towards the bottom of the cutting
chamber.
Turn the cooling valve to cleaning position (see sketch in
table) and adjust the flow as needed.
Clean the cutting chamber thoroughly.
NB! Do not clean the protection guard with the cleaning hose;
Turn the cooling valve to cooling position.
Leave the protection guard open to let the cutting chamber
dry completely.
Leave the protection guard open to let
the cutting chamber dry completely.
use a damp cloth instead.
AVOID RUST!