IEI
9.3.1. How to program particular pump heads (figs. 48,49)
If you have purchased a PD5201 or PD5206 in a product package, the drive has
already been adjusted to the respective pump head at the factory!
Turn power switch (29) OFF. Hit „Cal" key (34) and turn power switch (29) ON. Pump head
-
actually programmed is displayed.
Select different pump head with arrow keys (35), e.g. „Pu 2".
-
Confirm choice with „Select" key (36).
-
For coding of pump heads refer to reference table below.
-
code
Pu 0
Pu 1
Pu 2
Pu 3
Pu 4
Pu 8
Pu 9
9.3.2. Select display mode
Select key (36) allows to choose from several display modes:
speed rpm (speed in 1/min)
-
tube mm (hose ID in mm)
-
flow ml/min (feed rate in ml/min)
-
vol ml (metered volume in ml)
-
pause sec/min/h (pause in seconds, minutes, or hours) (interval metering)
-
9.3.3. Display feed rate in ml/min
In order to display feed rate in ml/min, the item needs to "know" about de hose diameter
-
used.
Feed rate is proportional with hose ID and pump speed.
-
Enter hose ID: hit select-key (36) several times until „Tube mm" LED lights.
-
Enter hose ID with arrow keys (35) (e.g. 0.8; 1.7; 3.1; 4.8; 6.3 or 7.9).
-
If the LED Vol ml blinks when you adjust the internal diameter of the hose, the
metering volume is being adapted to the minimum allowable volume. (See Chap.
9.3.5)
Hit select key (36) until LED „Flow ml/min" LED lights.'
-
Feed rate in ml/min is displayed (41).
-
To alter feed rate, use arrow keys (35) after having selected the "flow ml/min" mode with the
-
select key.
At this point, the processor is going to determine rotational speed.
-
pump head
multi-channel pump head C 8
SP quick / SP quick d
SP Standard / SP vario
SP mini (obsolete)
Masterflex
multi-channel pump head C 12
multi-channel pump head C 4
PD 5206
X
X
X
X
62
PD 5201
X
X
X
X
X
X
X