Mechanical Connections; Pump Location; Rotation Of The Pump Inlet With Respect To The Motor Frame; Pump Inlet - Agilent Technologies TriScroll 600 Série Manuel D'installation Et D'opération

Pompe à vide sèche
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17

Mechanical Connections

Pump Location

Locate the pump on a firm, level surface.
Mounting holes provided in the frame can be used to
provide permanent attachment. The rubber mounts on
the frame can be removed if desired.
CAUTION
The pump is a heavy object. To avoid
muscle strain or back injury, use lifting
aides and proper lifting techniques
when removing or replacing.
Rotation of the Pump Inlet
with Respect to the Motor Frame
The standard pump configuration is with the inlet fitting
positioned at the top of the pump as shown in Figure 1 on
page 35. The pump inlet has two alternate positions:
90 degrees clockwise
90 degrees counterclockwise
To rotate the inlet:
1. Remove three (3) M5 screws (item 1 on Figure 3 on
page 36) that secure the cowling (item 2 on Figure 3)
to the scroll module. Remove the cowling.
2. Tilt the pump back so that the rear end of the motor
touches the floor.
3. Remove the four (4) M6 screws (item 8 on Figure 3)
holding the frame (item 7 on Figure 3) and scroll
module together.
4. Axially separate the frame and scroll module.
5. Rotate the module 90 degrees in either direction and
realign the two (2) locating pins on the scroll module
with the mating frame holes. Ensure that the rubber
spider is still on the motor coupling and that the
coupling teeth are properly aligned.
6. Rejoin the scroll module with the frame and install
and tighten the four (4) M6 screws removed in step 3.
7. Reinstall the cowling using the three (3) M5 screws
removed in step 1.

Pump Inlet

Use NW40, or larger, clean vacuum hardware with as
short a length as practical between the pump and the
vacuum chamber.
Use a bellows to provide both vibration isolation and
strain relief between the pump and the vacuum chamber.
TriScroll 600 Series Vacuum Pump

Pump Exhaust

A female 1/4" National Pipe Thread exhaust fitting is
located underneath the scroll module. This fitting swivels
360 degrees. Additionally, an NW25 male adapter with
3/8" National Pipe Thread is provided.
To avoid overheating the pump, do not restrict the
exhaust flow with long lengths of small diameter tubing.
Use as short as practical lengths of NW25 diameter, or
larger, hardware.

Optional Isolation Valve

Scroll pumps return to atmospheric pressure quickly
when shut off, thus the installation of a fast acting,
automatic, normally closed isolation valve is strongly
recommended to prevent pump debris from being
transported back into the vacuum chamber when the
pump is turned off.
The opening of this valve should occur
simultaneously with or after pump startup.
Valve closing should occur before,
simultaneously with, but no later than 250 ms
after pump shut off.
Use an NW40 valve or larger and mount it as close as
possible to the pump inlet. Mounting to the pump inlet is
ideal. (See Figure 6).
.
Figure 6 Isolation Valve Location
Agilent offers a variety of manual, electromagnetic and
electropneumatic controlled vacuum valves for vacuum
applications. The Agilent Vacuum Pump Isolation (VPI)
Valve is highly recommended for vacuum pump isolation
applications. The VPI valve application and installation
information for use with TriScroll vacuum pumps has
been included below.
Agilent also sells the TriScroll 600 pump with the VPI
valve preinstalled. The part number is PTS06001UVPI.
Isolation Valve
Pump Inlet
39

Publicité

Table des Matières
loading

Table des Matières