17
Technical Information
Window Protocol
N.
Read/
Data
Write
Type
108
R/W
N
109
W
L
110
R/W
L
111
R/W
L
117
R/W
N
120
R/W
N
122
R/W
L
125
R/W
L
126
R/W
N
147
R/W
N
155
R
N
298/314
Description
Baud rate
Pump life/ cycle time/ cycle number reset
Interlock type (default = 1)
Analog output type: output voltage signal
proportional to frequency or power
Low Speed frequency (Hz)
Rotational frequency setting (Hz)
Set vent valve on/off
Set the vent valve operation
Vent valve opening delay (expressed in 0.2
sec)
Vent open time
See "vent connector" paragraph
Power limit applied
Read the maximum allowable power
TPS-compact User Manual / 87-901-010-01
Admitted Values
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 1)
2 = Pump temperature
3 = Pressure
1100 to "Maximum rotational
frequency"
(win 120)
(default = 1100)
For 74 FS
For 304 FS
1100 to 1167
900 to 1010
(default =
(default =
1167)
1010)
0 = Closed
1 = Open
(default = 1)
Automatic = 0 (see note 1.)
On command = 1 (see note
2.)
0 to 65535 (corresponding to
0 to 13107 sec)
(default = 15)
0 = infinite
1 bit = 0.2 sec
1 to 65535
watt
(H)