Troubleshooting Chart

Problem
Possible Causes
TMP doesn't
Scroll Pump motor thermal
reach NORMAL
protector open
OPERATION
Vacuum system leak
System will not
Excessive voltage drop
start
Fuses burnt-out
Poor ultimate
Vacuum system leak
pressure
Water in TPS-compact
Solvent in pump
Scroll Pump seals worn out
Poor conductance to TPS-compact
Hammering
Scroll pump overheated
noise
Debris in Scroll pump
Internal Scroll
Motor thermal protector is cycling
Pump runs
open and closed.*
intermittently
* The internal Scroll Pump is equipped with an auto-reset thermal motor protector.
This protector automatically shuts down the pump when it detects an overload
condition, and automatically restarts the pump when the motor has cooled to
within an acceptable temperature range.
306/314
Corrective Actions
Allow motor to cool. Identify cause of overload.
Locate and repair leak.
Check size and length of cable.
Check fuses and if necessary replace.
Locate and repair leak.
Flush the system with air or dry nitrogen.
Flush the system with air or dry nitrogen. Install trap
or filter.
Replace tip seals.
Replumb with shorter and/or larger diameter tubing.
Check ventilation to TPS-compact. Check ambient
temperature.
Flush the TPS-compact. Disassemble pump and
inspect.
Allow the scroll motor to cool. Identify the cause of
the overload.