8 TROUBLESHOOTING

For any further information and description, please refer to EVOLUTION 2, 3 or 3E console user manual
(#65016, 65038, 65051, 65060).
9 CARE & MAINTENANCE
The recommendations for cleaning, disinfection and sterilization are updated and available on MORIA's
website: http://www.moria-surgical.com.

9.1 INTRODUCTION

In case of unusual vibrations or noises, do not use the unit and contact your distributor.
Please contact MORIA for any other servicing.
The following are only recommendations. They must be adapted according to the laws in the country
of utilization. All the products and solutions for cleaning and disinfection must be used according to the
manufacturer's instructions.
MORIA recommends that you:
avoid touching potentially contaminated areas,
use gloves when performing cleaning and decontamination tasks.

9.2 INITIAL TREATMENT AND STORAGE

To avoid risks of condensation inside the packaging, unpack and store the devices in a clean, dry environment.
Do not store the instruments in an environment or next to other products that may possibly have a corrosive
or magnetic effect. Avoid any contact between devices, especially those comprised of different materials.
The device should be damage free and have no scratches or other surface defects.
Fragile devices or those requiring particular manipulation must be handled separately, taking special care to
protect delicate parts.
9.3 CLEANING-DISINFECTION- STORAGE OF THE HANDPIECE AND ELECTRICAL CABLE
Steps
What to do
1
Cleaning
• Dip the blade oscillation shaft of the motor in
a sterile distilled water bath and run the motor
back and forth 10 times.
• Remove the motor from the cup, unplug it from
the console, and wipe the oscillation shaft of the
motor with a lint-free cloth.
• Cable can be cleant with a lint-free cloth
moistened with cleaning solutions.
• Air-dry the shaft of the motor with clean, filtered
compressed air (medical compressed air).
Disinfecting & Drying
2
• Wipe each part of the motor with a lint-free cloth
moistened with disinfecting solutions.
• Use appropriate disinfecting solutions (spray
or wipe clean disinfectant) according to the
manufacturers' instructions.
• Dry carefully with disposable instrument cleaner
(lint-free) then with clean, filtered, compressed air.
Storage
3
10
MORIA
#65013-F-02.2017
IMPORTANT WARNINGS
• The handpiece must be carefully
cleaned after each surgical procedure.
• Do not use abrasives or scrapers to clean
any elements of the system. Doing so may
result in reduced precision, burrs and/or
irregular tissue resection.
• Do not immerse the motor in the sterile
distilled water bath.
• To prevent damaging the connectors and
to ensure proper functioning of the motor,
never pull on the cables and never hold
the motor by its cable.
• It is imperative to verify that the devices
are perfectly clean and free of any organic
or other residues.
• Do not gas sterilize (ETO) the motor.
• Do not autoclave the motor.
• No current sterilization process is
compatible with MORIA motors
• Failure to completely dry the inside of the
motor can result in oxidation.
• Do not store products that are not
completely dry; doing so may create rust
and result in irregular corneal resection.
• When not in use, the M2 system should be
kept in its storage box (#22519331) in a
dry atmosphere.