Technical Specification

Tab. 5 Technical Specification
Pumping speed (♦)
N
720 l/s
2
He
660 l/s
H
485 l/s
2
Ar
690 l/s
Max gas throughput (**) (■)
(with recommended forepump)
N
2
Ar
Compression ratio (■)
N
>1E+11
2
He
2E+08
H
3E+06
2
Ar
>1E+11
Foreline Tolerance (*)
N
10 mbar
2
He
10 mbar
H
> 4 mbar
2
Ar
8.5 mbar
Base pressure with recommended forepump
Max gas throughput:
(**)
• water cooling (ambient temperature = 25 °C; water temperature = 15 °C; backing pump
11.6 m
/h)
3
• air cooling (ambient temperature = 25 °C; backing pump 11.6 m
Foreline tolerance defined to satisfy following criteria:
(*)
• the Turbopump still produces a compression of 100
• the rotational frequency is not below the low speed of the pump (values measured in water
cooling mode)
TwisTorr 804 FS User Manual / 87-901-052-01
Air cooling
4.3 mbar l/s – 255 SCCM
1.5 mbar l/s – 88 SCCM
< 3.5 x 10
mbar
-10
(< 2.6 x 10
Torr)
-10
(According to standard DIN 28 428, the base pressure is
that measured in a leak-free test dome, 48 hours after the
completion of test dome bake-out, with a Turbopump fitted
with a ConFlat flange and using the recommended pre-
vacuum pump)
Technical Information