Installation

WARNING!
The power cord of the controller is defined as disconnecting device in order to
remove the power from the pump and the contoller in case of an emergency.
Make sure the power connector of the instrument can be easily reached and
unplugged. Provide sufficient space behind the power socket of the controller
to unplug the cable.
If the controller cannot be installed in the system, so that the power cord can
be easily reached, a different option has to be installed to disconnect the
controller from mains.
Do not remove the adhesive and protective cap before connecting the turbopump
CAUTION!
to the system.
Figure 2
Install this device indoors only, and in any case do not use the pump
in an environment exposed to atmospheric agents (rain, ice, snow),
dust, aggressive gases, or in explosive environments or those with a
high fire risk.
During operation, the following environmental conditions must be
respected:
TwisTorr 804 FS User Manual / 87-901-052-01
maximum pressure: 2 bar above atmospheric pressure;
temperature: from +5 °C to +35 °C;
Instructions for Use

Installation

263/344