Télécharger Imprimer la page

Functions - Siemens SIPAN 32 Instructions De Service

Masquer les pouces Voir aussi pour SIPAN 32:

Publicité

Les langues disponibles

Les langues disponibles

Mise en service et fonctionnement
4.8.7.2.2
4-50
MPIP sarl - 94, rue Mohamed Tounsi - 16013 - ROUIBA - ALGERIE
Tél./Fax : 021 81 11 22 - GSM : 070 91 75 15 ou 072 900 654 - E-mail : mpip@mpi-dz.com

Functions

The setting possibilities via SIMATIC PDM and the views of the results
are listed below:
Identification
Specifications on operating reliability:
S
day, description and message
Specifications for the device, among others:
S
Product name, serial number and ordering information, SW and HW
revision
Input
Specifications on the measuring method:
S
calibration factor, unit, component name, scanning interval,
measuring range
Output
Output calibration:
S
Redefinition of the output range for the application (related to the
unit in the output signal)
Output signal:
S
Redefinition of the unit for the application, furthermore filter time
constant for attenuation of the interference suppression,
parameterizable failure behavior, e. g. set equivalent value.
Output limit values:
S
Setting the alarm and warning limits
Settings by menus or methods
Simulation: Here the measured value (primary variable) or the whole
S
device including output section can be simulated.
Certificates and licenses
Setting whether or not the transmitter is to be operated in the
S
intrinsically safe range.
View of the measured values and diagnostic
Register for measured values:
S
Primary variable (main measured value) related to the parameter
settings which have been described in the section "Input" (see
above).
Register for output values:
S
Converted primary variable (main measured value and
channel--related measured values) as described in section "Output".
Measured value calibration and output calibration have been
performed. The output value with status is sent cyclically to the
process control system.