4.8.4

Mode of Operation

Dispositif de mesure de conductivité SIPAN 32
C79000-B5477-C045-02
MPIP sarl - 94, rue Mohamed Tounsi - 16013 - ROUIBA - ALGERIE
Tél./Fax : 021 81 11 22 - GSM : 070 91 75 15 ou 072 900 654 - E-mail : mpip@mpi-dz.com
The SIPAN 32 PA transmitter family can be used in all branches.
The following sensors can be connected:
Conductivity sensors: 2EL, 4EL, IND
S
pH/ORP sensors
S
Sensors for dissolved oxygen
S
The useful data -- measured values with status as a quality indication --
are provided on PROFIBUS--PA (see section 4.8.8 "Useful Data").
SIPAN 32 X PA in the "intrinsically safe ignition protection type" can be
installed in hazardous areas (Ex--zone 1) and can be used for feeding
sensors at zone 0. The conformity declarations comply with the
European standard (CENELEC).
Product Features
SIPAN 32 PA with bus connection according to IEC 61158--2 and
S
EN 50170, Part 4
Data transmission and device supply via common bus connection
S
Communication via PROFIBUS--PA (Profile B, Version 3.0); sensor,
S
measuring range and much more can therefore be programmed.
Quality data for the measured values: status with limit values
S
Fixed bus current limiting in the event of an error
S
Electrical isolation (test voltage 500 V AC)
S
Intrinsically safe version for use in hazardous areas
S
Mise en service et fonctionnement