Burkert 8205 Notice D'utilisation page 60

Masquer les pouces Voir aussi pour 8205:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3 INSTALLATION
PH CONTROLLER 8205
3.1.2 pH controller 8205 panel installation
Respect the installation specifications described in the instruction manual of the remote
sensor.
Follow the instructions below to install the transmitter:
1. Use the supplied drilling jig to cut away the necessary opening in the panel. Make sure
to scrupulously respect the dimensions indicated.
2. Place the gasket on the 4 screws of the cover.
Note: Use the 4 M4x25 screws supplied if the panel door is too thick.
3. Place the cover+gasket set on the cut-away with the electronic board facing the inside
of the panel.
4. Insert the 4 washers onto the 4 screws and fit the device to the panel using the 4 nuts.
5. Connect according to instructions described in chapter 3.2.
6. Use the supplied cable clip to fit the cable to the protective plate (see 3.2.2).
Screw
Washer
Nut
Gasket
Cable clip
Fig. 3.2
Mounting pH controller, panel-mounted version
E-10-

Publicité

Table des Matières
loading

Table des Matières