Instructions for use: rigid medical endoscopes

11. Disassembly

WARNING: Risk of burns!
Allow the illumination fibre to cool sufficiently before removing it. The ends can become very hot and cause
serious burns.
Remove illumination fibre.
Do not remove the ocular funnel because otherwise the endoscope will be damaged.
Unscrew existing adapters.
For endoscopes with working channel, disassemble the adapter with stopcock.
- Remove sealing cap c.
- Loosen union nut a.
- Loosen stopcock nut d.
- Disassemble stopcock b.

12. Storage

Unsterile metal units must be stored in a clean, dry environment. The storage time of unsterile units is not limited;
the units are made of a non-degradable material which maintains its stability when stored under the recom-
mended conditions.
As long as endoscopes are stored unsterile in the original packaging, the following storage conditions apply:
- Temperature: -10°C to +40°C,
- Humidity: 10% to 90%.
Avoid direct sunlight.
Store endoscope either in the original packaging or in a screen tray/container.
Ensure that the endoscope is stored securely.
Observe the respective valid national provisions when storing in a sterile condition.

13. Service and maintenance

Delmont imaging does not supply original parts to independent workshops or other endoscope manufacturers.
Thus only Delmont imaging is in a position to carry out repairs using original parts. The original technical specifi-
cations and the operational safety of the endoscope can only be guaranteed by using original parts.The warranty
for Delmont imaging products shall become void if repairs are carried out by an workshop not authorised by
Delmont imaging. In this case Delmont imaging is also no longer responsible for the technical specifications or
safety of the product.
Have the endoscopes repaired by Delmont imaging only. For service, send the defective endoscope to the
address of the sales partner.
Clean, disinfect and sterilise the endoscope thoroughly prior to returning it for repair.
Ideally, send in the endoscope in its original packaging. If this is not possible, package the endoscope to
secure it for transport.
Delmont imaging is not liable for damage resulting from improper shipping.
D900 700 023 D
Rev 10/2018
delmont
imaging