3. For the AC model: Unplug the incoming AC power connector from the IND246
main board, note the orientation of the black and white wires, loosen the two
small screws fastening the wires in place and detach the wires.
4. For the battery-powered model: Unplug the power connector from the main
board (Figure 2-5).
5. Loosen and remove the two nuts securing the two hinge/grounding straps that
connect the front panel to the rear housing.
6. For the AC model: Cut the nylon wire tie securing the ferrite core on the AC
wires to the plastic adhesive pad.
7. Note how the two AC power wires are looped through the ferrite core and then
remove the ferrite core from the two wires.
8. Unscrew the nut that secures the green/yellow ground wire to the enclosure
and remove the wire and loop terminal from the threaded stud.
9. Loosen and remove the cable gland and AC power cable from the rear of the
enclosure.
10. For both models: Remove the load cell cable gland from the rear of the
enclosure and reinstall it into the same sized opening at the other side of the
enclosure (where the AC cable gland was located in the AC model). Tighten
the gland to 2 N-m (18 lbf-in).
11. For the AC model: Install the AC power cable and gland into the open location
where the load cell cable gland was previously installed. Tighten the gland to
2 N-m (18 lbf-in).
12. Use the nut removed in step 8 above to secure the green/yellow ground wire to
the stud as shown.
13. Loop the two AC wires through the ferrite core in the same manner noted in
step 7 and secure the ferrite to the plastic adhesive pad with the new wire tie
provided.
14. For both models: Carefully rotate the housing 180 degrees and reattach the
two grounding straps to the two studs near the grip bushings (Figure 2-6)
using the two nuts removed in step 5. Tighten the two nuts.