9.3 Input Data
Address without offset
0
4
8
The data must be processed (copied) into especially created tags in a correct format in accordance to column
"Format". For example, to obtain the internal temperature of the device, one should create a REAL tag and an
instruction copying 4 bytes of the device's input data into this tag, beginning with byte 12.
9.4 Output Data
Once the pyrometer has been initialized and is running in the data exchange mode, only the below listed
parameters can be changed, using the device's output data.
Address without offset
0
1
The <Type of parameter> gives the meaning of the following parameters with the same format as described in
section 9.2 Parameters, page 65.
66
1.800.561.8187
Table 9-13: Pyrometer Parameters
Attribute ID
Name
Data type
0x01
Temperature unit
0x02
Emissivity
0x03
Transmissivity
0x04
Device Offset
0x05
Device Gain
0x06
Average Time
0x07
Peak hold Time
0x08
Valley hold Time
0x09
Laser control
Table 9-14: Input Data
Length
Format
4 Byte
REAL (Big Endian, Motorola)
4 Byte
REAL (Big Endian, Motorola)
4 Byte
DWORD
Table 9-15: Output Data
Length
Format
1 Byte
BYTE
4 Byte
REAL / UDINT
www.
Data value
USINT
0x43 ('C') – Celsius
0x46 ('F') - Fahrenheit
REAL
0.1 ... 1.1
REAL
0.1 ... 1.1
REAL
-200 ... 200
REAL
0.8 ... 1.2
REAL
0.1 ... 999.0
REAL
0.0 ... 999.0
REAL
0.0 ... 999.0
USINT
0 - off, 1 - on
Value
Target temperature
Internal temperature
Error code
Value
Type of parameter
Parameter
information@itm.com
.com
Access rule
Get/Set
Get/Set
Get/Set
Get/Set
Get/Set
Get/Set
Get/Set
Get/Set
Get/Set