The <Type of parameter> gives the meaning of the following parameters with the same format as described in
section 8.2 Parameters, page 62.
If <Type of parameter> is set to 0 then the output data gets ignored. As default, it should be set to 0 (zero).
8.6 Diagnostics
The diagnostics information of the fieldbus communicator can be read out acyclically using standard diagnostics
data sets defined in the PROFINET IO specification.
Errors occurring when configuring and setting the parameters of the fieldbus communicator and the connected
pyrometer as well as external errors are reported by the communicator via channel specific diagnostic.
In productive data exchange between the I/O controller and the fieldbus PROFINET IO, one byte IOPS process
data qualifiers are available for each module providing information of the validity of the pyrometer module data
(good/ bad). In the event of an error occurs during operation, the problem-indicator in APDU-Status is set by the
communicator and a diagnostic alarm is additionally transmitted.
64
1.800.561.8187

Table 8-11: Parameter Types

Type of parameter
Meaning
0
Do not change anything
1
Emissivity
2
Transmissivity
3
Averaging time
4
Peak hold time
5
Valley hold time
6
Laser control

Table 8-12: Error Codes

Bit
Description
2
Internal temperature over range
3
Internal temperature under range
9
Object temperature under range
10
Object temperature over range
17
PROFINET not ready