Messages; Nput Data; Output Data; Table 8-7: Pyrometer Parameters - Fluke Thermalert 4.0 Manuel Utilisateur

Table des Matières

Publicité

Temperature unit
Emissivity
Transmissivity
Average Time
Peak hold Time
Valley hold Time
Laser control
Gain
Offset

8.3 Messages

Message diagnostics alarm
Message process alarm
Behavior on module fault
8.4 Input Data
Address without offset
0
4
8

8.5 Output Data

The output data may be used to change the initialization of the device (which was set once at start-up) when the
bus is in data exchange mode.
Address without offset
0
1
1.800.561.8187

Table 8-7: Pyrometer Parameters

Parameter
Set the temperature unit
*1000, in thousandths (0.9 → 900)
*1000, in thousandths (1.0 → 1000)
*0.1 s, in tenths (1 s → 10)
*0.1 s, in tenths (1 s → 10)
*0.1 s, in tenths (1 s → 10)
Amplification adjustment for temperature
value
Shift adjustment for temperature value
Parameter
The diagnostics information of pyrometer is not
transferred to the PROFINET IO controller
is transferred to the PROFINET IO controller
The process alarm of pyrometer is not transferred
to the PROFINET IO controller
is transferred to the PROFINET IO controller
Length
4 Byte
4 Byte
4 Byte
Length
1 Byte
4 Byte
www.
Description

Table 8-8: Messages

Description

Table 8-9: Input Data

Format
REAL (Big Endian, Motorola)
REAL (Big Endian, Motorola)
DWORD

Table 8-10: Output Data

Format
BYTE
REAL / UDINT
.com
Setting
Celsius, Fahrenheit
100...1100
100...1100
1 ... 9990
0 ... 9990
0 ... 9990
off/on
0.8000 ~ 1.2000
-200.0 ~ 200.0°C / -360.0 ~ 360.0°F
In current unit (°C/°F)
Setting
message inactive (default)
message active
message inactive (default)
message active
set process data to zero, set process data
to last value
Value
Target temperature
Internal temperature
Error code
Value
Type of parameter
Parameter
information@itm.com
8
PROFINET IO
Messages
63