Pressure Display Module - Agilent Technologies TPS-mini Notice De Mode D'emploi

Pompe turbomoléculaires
Masquer les pouces Voir aussi pour TPS-mini:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 45
Pressure gauge can be connected/disconnected from the TPS-mini
during normal operation.
Figure 18
The FRG700 gauge is intended for high vacuum use, so with pressure above 50
NOTE
mTorr the signal is misinterpreted and a "false" pressure is displayed. Use a
transition gauge to understand if the reading is really under 50 mTorr or not.

Pressure Display Module

Figure 19
TPS-mini User Manual / 87-901-003-01 (E)
Gauge Connector
Pressure display module
Technical Information
P1 – Vent
285/318
17

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969918096991819699182

Table des Matières