Agilent Technologies TPS-mini Notice De Mode D'emploi page 302

Pompe turbomoléculaires
Masquer les pouces Voir aussi pour TPS-mini:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 45
17
Technical Information
Window Protocol
Tab. 6
Data Type
Logic (L)
Numeric (N)
Alphanumeric (A)
The addressed SLAVE will respond with an ANSWER whose structure
depends from the MESSAGE type.
When the MESSAGE is a reading command, the SLAVE will respond
transmitting a string with the same structure of the MESSAGE.
Using the RS 485 interface, the message structure remains identical to the one
NOTE
used for the RS 232 interface, the only difference being that the value assigned
to the ADDRESS <ADDR>.
302/318
Field Length
1
6
10
<ETX> (End of transmission) = 0x03
<CRC> = XOR of all characters subsequent to <STX> and
including the <ETX> terminator. The value is hexadecimal coded
and indicated by two ASCII character.
Valid Characters
'0' = OFF
'1' = ON
'-', '.', '0' . . . '9' right justified with '0'
from blank to '_' (ASCII)
TPS-mini User Manual / 87-901-003-01 (E)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969918096991819699182

Table des Matières