Controller Description; Diaphragm Forepump Description; Control Panel Description - Agilent Technologies TPS-mini Notice De Mode D'emploi

Pompe turbomoléculaires
Masquer les pouces Voir aussi pour TPS-mini:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 45

Controller Description

The integrated controller is a solid-state frequency converter which is
driven by a single chip microcomputer and is composed of a PCB
which includes a power supply with a 3-phase AC output, analog and
input/output section, microprocessor and digital section. The
controller recognizes the mains presence and converts the single
phase AC mains supply into a 3-phase, medium frequency output
which is required to power the pump.
The controller can be operated by a remote host computer via the
serial connection.

Diaphragm Forepump Description

The pump consists of an electric DC motor driving an eccentric
which moves an elastic diaphragm. The pump has four heads.The
compression chamber is hermetically separated from the drive
mechanism by means of the diaphragm.
The pump transfers, evacuates and compresses completely oil-free.
The motor is supplied with a voltage of 24 Vdc.
A diaphragm pump cleaning from the humidity is recommended to improve the
NOTE
system performance using dry air or nitrogen. It is recommended to open
manually the vent valve after every start command for about 30 seconds.

Control Panel Description

The following picture shows the TPS-mini control panel; the
associated table lists the available commands. For a detailed
description see the following paragraphs.
TPS-mini User Manual / 87-901-003-01 (E)
Technical Information
Turbopump Description
271/318
17

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969918096991819699182

Table des Matières