MSA Orion G Mode D'emploi page 70

Masquer les pouces Voir aussi pour Orion G:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Chapter 7, Performance Specifications
Table 7-2. Instrument Specifications
Normal
TEMPERATURE
RANGE
Extended**
WARM-UP TIME
*NOTE 1: Response times are for the instrument without sample lines.
When using a sample line, add time for the gas sample to be drawn through the
sample line. Typical transport times are:
5 feet
10 feet
50 feet
These should be added to the sensor response times reported in this chapter.
**NOTE 2: Extended temperature range indicates that gas readings may vary
slightly if calibrated at room temperature. For optimal performance, it is
recommended that the instrument be calibrated at temperature of use.
COMBUSTIBLE
Catalytic Sensor
GASES
OXYGEN
Electrochemical Sensor
TOXIC GASES
Electrochemical Sensors
CH
LEAK
Metal oxide semiconductor
4
HIGH ALARM LOW ALARM
CO
35 PPM
H
S
10 PPM
2
COMBUSTIBLE
0.5 %
- CH
(10% LEL)
4
O
22.0%
2
LEAK
--
7-2
o
0 to 40
C
o
-20 to 50
C
20 seconds; 25 seconds with pump
4 seconds
8 seconds
40 seconds
Measurement Methods
Factory-Set Alarm Setpoints
--
--
--
19.5%
--
Operating Manual
STEL
TWA
400
35
15
10
--
--
--
--
--
--

Publicité

Table des Matières
loading

Table des Matières