Cleaning Instructions

Disinfection of contaminated parts is only effective after having them
carefully pre-cleaned. CareFusion recommend the tested solution of
PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer's instructions.
The device must not be exposed to solvents e.g. alcohol, chloride
CAUTION: Unplug your SpiroUSB before cleaning.

External Surfaces of the Spirometer

CAUTION: Do not attempt to wash or immerse the SpiroUSB
transducer housing in water or cleaning fluid, as there are electronic
components inside that will be permanently damaged
The external housing of the spirometer may be wiped with sterile
wipes or a damp cloth that has been immersed in a cold sterilising
solution when required.

Cleaning accessories

With the use of a filter MicroGard® IIB (V-892381)
MicroGard® IIC
,
(V-892384) or a One-way Safety mouthpiece (36-SST1250) for each
patient, cleaning for the components in patient's gas path is
recommended once a month.
When using the disposable cardboard mouthpiece (adult: 36-
PSA1000, pediatric: 36-PSA1200) without a filter and under the
prerequisite that the patient was instructed only to exhale into the
transducer, the following parts have to be cleaned once a day:
adult/ pediatric adapter, transducer.
With any other use as described above all contaminated parts must
be disinfected between patients.