CAUTION: Read the manual before use
WARNING: The instrument is not suitable for use in the presence
of explosive or flammable gases, flammable anaesthetic mixtures or
in oxygen rich environments.
CAUTION: Mouthpieces are single patient use. If used on more than
one patient there is a risk of cross-infection. Repeat use may
degrade materials and lead to an incorrect measurement."
CAUTION: Pulmonary filters are single patient use. If used on more
than one patient there is a risk of cross-infection. Repeat use may
increase air resistance and lead to an incorrect measurement."
PLEASE NOTE:
not be disposed of as unsorted waste. Please utilise your
local WEEE collection facilities for the disposal of this
product.
PLEASE NOTE: Degree of protection against Ingress of Water is
IPX0
WARNING: To avoid risk of electric shock, this equipment must
only be connected to a supply mains with protective earth
CAUTION: When you connect the SpiroUSB to other equipment,
always make sure that the whole combination complies with the
international safety standard IEC 60601-1 for medical electrical
systems. During measurements, connect the SpiroUSB only to
printers and computers that comply with IEC/EN 60601-1 / UL
60601-1.
Indication for Use
The SpiroUSB spirometer is intended, for prescription use only, to
measure the maximal volume and flow of air that can be moved in
and out of a patient's lungs. The system is intended for use with
paediatric (4 to 17 years) and adult (18 to 99 years) patients in
hospitals, physician offices, laboratories and occupational health
testing environments.
The product you have purchased should
16