Cleaning The Transducer - CareFusion SpiroUSB Manuel D'utilisation

Masquer les pouces Voir aussi pour SpiroUSB:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 71

Cleaning the Transducer

The transducer requires no routine maintenance or servicing.
However, if you wish to sterilize or clean the transducer it may be
removed by means of the following procedure:
1. Rotating the turbine transducer anti-clockwise until the
locating pip lines up with the small rectangular cut-out in the
housing as shown below.
2. Gently pull the transducer away from the housing.
3. The transducer may now be immersed in warm soapy water
for routine cleaning or immersed in a cold sterilizing solution
e.g. Perasafe (Cat No. 36-SSC5000A) for a maximum of 10
minutes (Alcohol and chloride solutions should be avoided).
4. After cleaning/sterilizing, the transducer should be rinsed
briefly in distilled water and dried.
5. Re-assemble the mouthpiece holder.
Turbine transducer
CAUTION: Do not attempt to wash or immerse the SpiroUSB
transducer housing in water or cleaning fluid, as there are electronic
components inside that will be permanently damaged
Housing
Locating Pip
50
Cut-Out

Publicité

Table des Matières
loading

Table des Matières