Combinations in Long pulse width mode:
Watts
5 Hz
0.5 J
2.5
0.6 J
3.0
0.7 J
3.5
0.8 J
4.0
1.0 J
5.0
1.2 J
6.0
1.5 J
7.5
2.0 J
10.0
2.5 J
12.5
3.0 J
15.0
Turning the System Off
To turn off the H-30, place the system into Standby mode, turn the key switch to the off
(O) position and remove the key.
WARNING! If the system is turned off without entering Stand-
by mode, the parameters will not be saved.
Protection Against Unauthorized Use
When the H-30 is not in use, protect it against unauthorized use by removing the key from
the key switch.
Error Messages
There are several error messages that can be displayed either in the information bar on
the main screen or in the error log on the options screen.
The following are possible types of messages:
• Simmer failure: The flash lamp doesn't simmer. Call for service.
• Shutter failure: The shutter isn't functioning properly. Call for service.
• Over temperature: The temperature of the cooling water is too high. Call for service.
• Low coolant flow: If water level is low, fill the reservoir with distilled water.
Cook Medical © 2012
8 Hz
10 Hz
12 Hz
4.0
5.0
4.8
6.0
5.6
7.0
6.4
8.0
8.0
10.0
12.0
9.6
12.0
14.4
12.0
15.0
18.0
16.0
20.0
24.0
20.0
25.0
Return to Table of Contents
15 Hz
20 Hz
6.0
7.5
10.0
7.2
9.0
8.4
10.5
9.6
12.0
15.0
18.0
22.5
Chapter 5
Operation
1-29