Backlighting Of The Display; Centrifuge Address - Hettich ROTANTA 460 Robotic Mode D'emploi

Masquer les pouces Voir aussi pour ROTANTA 460 Robotic:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

30

Backlighting of the display

(only possible starting from program version V 01.18)
To save energy, it can be set that, after a centrifugation run, the backlighting of the display switches off after 2
minutes.
With the rotor at a standstill, this can be set as follows:
This setting can only be made via the operating unit (key position "0" or "LOCK 2"). The functions of the
various key positions are described in the "Key-operated switch" chapter.
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
Press the
key as often as necessary until the ->Settings is displayed.
PROG
Press the
key. The SOUND / BELL = off/on is displayed.
START
Press the
key as often as is required until Power save = off/on is displayed.
PROG
Power save : Automatic shutdown of the backlighting.
Use the adjusting knob to set
off = Automatic shutdown deactivated,
on = Automatic shutdown activated.
Press the
key to save the setting.
START
Briefly, Store Settings .. is displayed as confirmation, followed by -> Settings .
Press the
key once to exit the "Settings" menu or press the
OPEN / STOP
"Machine Menu".
31

Centrifuge address

Ex works, the address is ] = 29. Address is set.
AB5680DEENFRIT
off or on .
Rev. 12 / 04.2020
key.
T/°C
key. In this case, the settings are not
OPEN / STOP
key twice to exit the
OPEN / STOP
EN
83/179

Publicité

Table des Matières
loading