30
Backlighting of the display
(only possible starting from program version V 01.18)
To save energy, it can be set that, after a centrifugation run, the backlighting of the display switches off after 2
minutes.
With the rotor at a standstill, this can be set as follows:
This setting can only be made via the operating unit (key position "0" or "LOCK 2"). The functions of the
various key positions are described in the "Key-operated switch" chapter.
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored
•
Keep the
key held down for 8 seconds.
PROG
After 8 seconds, ∗∗∗ Machine Menu ∗∗∗ appears in the display.
•
Press the
key as often as necessary until the ->Settings is displayed.
PROG
•
Press the
key. The SOUND / BELL = off/on is displayed.
START
•
Press the
key as often as is required until Power save = off/on is displayed.
PROG
Power save : Automatic shutdown of the backlighting.
•
Use the adjusting knob to set
off = Automatic shutdown deactivated,
on = Automatic shutdown activated.
•
Press the
key to save the setting.
START
Briefly, Store Settings .. is displayed as confirmation, followed by -> Settings .
•
Press the
key once to exit the "Settings" menu or press the
OPEN / STOP
"Machine Menu".
31
Centrifuge address
Ex works, the address is ] = 29. Address is set.
AB5680DEENFRIT
off or on .
Rev. 12 / 04.2020
key.
T/°C
key. In this case, the settings are not
OPEN / STOP
key twice to exit the
OPEN / STOP
EN