EN
1
Intended use
This device is a medical product (laboratory centrifuge) within the context of the IVD Directive 98/79/EC.
The centrifuge was designed only for the separation of materials or mixtures with a density of no more than 1.2
kg/dm³. In particular, these include samples for preparation of in-vitro diagnostic purposes in human medicine.
The centrifuge is only meant for this purpose.
Another use or one which goes beyond this, is considered to be non-intended. The company Andreas Hettich GmbH
& Co. KG is not liable for damage resulting from this.
Observing all information in the operating instructions and complying with the measures described therein is also a
part of the intended use.
If the centrifuge is installed in another device or in a system, the manufacturer of the overall system is responsible for
its safety.
2
Remaining risks
The device is built according to the state-of-the-art and the recognized safety regulations. If used and handled
improperly, there could be life-threatening danger to the user or third parties, or the device could be impaired or there
could be other property damage. The device is only to be used for its intended purpose and only when it is in safe
working condition.
Malfunctions which could affect safety must be corrected immediately.
3
Technical specifications
Manufacturer
Model
Type
Mains voltage ( ± 10%)
Mains frequency
Connected load
Cooling medium
Max. capacity
Allowed density
Speed (RPM)
Force (RCF)
Kinetic energy
Obligatory inspection (BGR 500)
Ambient conditions (EN / IEC 61010-1)
− Set-up site
− Altitude
− Ambient temperature
− Humidity
− Excess-voltage category
(IEC 60364-4-443)
− Pollution degree
Device protection class
EMC
− Emitted interference,
Interference immunity
Noise level (dependent on rotor)
Dimensions
− Width
− Depth
− Height
Weight
56/179
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
ROTANTA 460 Robotic
5680-70
5680-78
230-240 V 1 ∼
200-220 V 1 ∼
Up to 2000 m above sea level
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Not suitable for use in explosion-endangered areas.
EN / IEC 61326-1, Class B;
697 / 762* mm (* with cables connected)
Rev. 12 / 04.2020
5680-75
5680-71
127 V 1 ∼
110-120 V 1 ∼
50-60 Hz
max. 1800 VA
R452A
4 x 750 ml
3
1.2 kg/dm
6200
5845
40000 Nm
yes
Indoors only
5°C to 35°C
ΙΙ
2
Ι
FCC Class B
≤ 68 dB(A)
554 mm
693 mm
approx. 154 kg
AB5680DEENFRIT
5680-74
100 V 1 ∼