Enter Centrifugation Parameters; Running Time; Starting Runtime Timing; Speed (Rpm) - Hettich ROTANTA 460 Robotic Mode D'emploi

Masquer les pouces Voir aussi pour ROTANTA 460 Robotic:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

18

Enter centrifugation parameters

Parameters can be input via the operating unit (key position "0") or via the interface (key position "LOCK 2").
The functions of the various key positions are described in the "Key-operated switch" chapter.
If no key is pressed for 8 seconds long after the selection or during the input of parameters, the previous
values will be shown in the display. The input of parameter then has to be executed again.
If several parameters are input, the
accept all changes.
If parameters are changed, the program place number is displayed in parentheses (). This means that the
centrifugation data in the display no longer corresponds to the centrifugation data from the program place that
has been saved.
You can no discontinue entering parameters at any time by pressing the key
adjustments are not saved.
18.1

Running time

In order to set continual running, the minutes, seconds and hours have to be set to zero.
Continual running is represented in the display by the following symbol, "∞".
Press the
key. The parameters t/hms is displayed. The minutes ( m) are shown in parentheses 〈〉, and can
TIME
be changed.
Use the adjusting knob
Press the
key. The seconds ( s ) are shown in parentheses 〈〉and can be changed.
TIME
Use the adjusting knob
Press the
key. The hours ( h ) are shown in parentheses 〈〉 and can be changed.
TIME
Use the adjusting knob
To apply the setting to the display, either press the
the centrifugation data are displayed.
18.2

Starting runtime timing

The start to runtime timing can only be set if the function, "Dual time mode" has been activated. See chapter
"Activate or deactivate function "Dual time mode". This function is activated ex works.
Press the
key as often as is required until Timing begins at start or Timing begins at speed is displayed.
TIME
Select the desired setting with the turning knob .
Timing begins at Start
Timing begins at Speed = The runtime only begins to count after the set speed is reached.
Press the
or
TIME
START
18.3

Speed (RPM)

Press the
key. The parameters RPM is displayed.
RPM
Use the adjusting knob
Press the
or
RPM
START
AB5680DEENFRIT
key must be pressed after setting the last parameter in order to
START
to set the value you want.
to set the value you want.
to set the value you want.
START
= The runtime begins to count immediately after the centrifugation run starts.
This is indicated on the display by the symbol , to the left next to the time.
keys to apply the setting to the display.
to set the value you want.
keys to apply the setting to the display
Rev. 12 / 04.2020
OPEN / STOP
key or press the
key as often as is required until
TIME
EN
. In this case the
71/179

Publicité

Table des Matières
loading

Table des Matières