The following alarm parameters can be set up:
Temperature 1
ON
OFF
ON
Signal
OFF
40 [s]
Time OK
40 [s]
Time ERR
3,0 [°C]
min.
3,0 [°C]
max.
Door contact
ON
OFF
ON
Signal
OFF
5 [s]
Time OK
2 [s]
Time ERR
Speed
ON
OFF
ON
Signal
OFF
20 [s]
Time OK
20 [s]
Time ERR
10 [U/min]
min.
10 [U/min]
max.
Power failure
ON
OFF
ON
Signal
OFF
The temperature alarm is active
The temperature alarm is inactive
Acoustic signal active in case of an alarm
Acoustic signal inactive in case of an alarm
Duration of time, during which the set temperature must have been at least
reached after a temperature error, before the next error can be registered
(here 40 seconds).
Duration of time, during which the temperature error must at least occur,
before the error is indicated (here 40 seconds).
Permitted temperature deviation downwards, where no error is indicated
(here 3,0°C)
Permitted temperature deviation upwards, where no error is indicated (here
3,0°C)
The door contact alarm is active
The door contact alarm is inactive
Acoustic signal is active if the alarm occurs
Acoustic signal is inactive if the alarm occurs
Duration of time, during which the plexiglass cover must be at least closed,
before the error is registered again (here 5 seconds)
Duration of time, during which the plexiglass cover must be at least opened,
before the error is indicated (here 2 seconds)
The speed alarm is active
The speed alarm is inactive
Acoustic signal active at alarm message
Acoustic signal inactive at alarm message
Duration of time, during which the set speed after a speed error must have
been at least reached again, before the next error can be registered (here 20
seconds)
Duration of time, during which the speed error must at least occur, before
the error is indicated (here 20 seconds)
Permitted speed deviation downwards, where no error is indicated (here 10
U/min.)
Permitted speed deviation upwards, where no error is indicated (here 10
U/min.)
The power failure alarm is active
The power failure alarm is inactive
Acoustic signal is active if the alarm occurs
Acoustic signal is inactive if the alarm occurs
27