Checking the door contact switch
Whenever you start to operate the CERTOMAT ® Is, you
have to check the proper function of the door contact
switch!
To check the function of the switch, proceed as
follows:
After turning on the CERTOMAT ® IS and initialisation,
the unit is in the Manual-Menu „MAN".
•
In the manual mode „MAN" enter these values via the
keypad: (see also chapter 3.3.3)
1
F1
speed [
/min.]:
F3
time [h : min]:
•
Press the START-key to start the shaker.
Open the hood of the CERTOMAT ® IS.
The CERTOMAT ® IS must stop.
Once the preset time has expired (see chapter 3.3.5),
the alarm LED on the front panel lights up. If activated,
an alarm signal is generated in addition.
•
Press ENTER to acknowledge the alarm.
•
Close the hood of the CERTOMAT® IS. The shaker re-
starts.
If the shaker continues to run inspite of the openend
hood, the unit is defective.
In this case the CERTOMAT ® IS must not be started!
Repairs can be carried out by the Sartorius AG service
representative or by authorized personnel of a
customer-owned workshop. Please inform your local
Sartorius AG sales representative or contact:
Sartorius AG
Service-Zentrum Nord
Weender Landstraße 94-108
37075 Goettingen, Germany
Telephone +49.551.308.3729 / 3740 / 3741
Telefax +49.551.308.3730