8 Maintenance
60
8.4
Servicing Midwest quick coupling
8.4.1
Replacing the halogen lamp of the 6-pin coupling
CAUTION!  The lamp can be hot. Risk of burns! Allow the lamp to
cool down.
1.
Remove the high-speed handpiece.
A
2.
Insert blunt pick into recessed area on lamp base and slide lamp out
of socket.
3.
Remove the defective lamp (A).
IMPORTANT: Do not touch replacement lamp with bare fingers. Oils
from the skin decrease lamp life. Use gloves when handling lamp.
4.
Insert carefully the new lamp with contact pin end first into lamp
socket.
Be sure to orient recessed area on lamp base to face out slot on
coupling. If it is not possible to insert lamp, rotate lamp 180 degrees
and try again.
5.
Check the light function of the quick coupling.
Instructions for Use
66 07 290  D3673.201.01.08.02