4 Preparation
28
4.3
Before each patient
CAUTION! Insufficient cooling leads to overheating of the
preparation site and to damage to the tooth substance. Ensure that
the water content is ≥ 50 ml/min.
1.
Purge the water and air channels for 30 seconds.
2.
Attach the high-speed handpiece.
3.
Insert the bur.
4.
Set a sufficient amount of cooling water (> 50 ml/min) [→ 36].
5.
Use filtered water only (< 80 µm).
6.
Check the nozzles for blockages and lime deposits, for example,
and clean the nozzles if necessary.
Instructions for Use
66 07 290 D3673.201.01.08.02