Instructions for Use
66 07 290  D3673.201.01.08.02
5
Operation
CAUTION!  Insufficient cooling leads to overheating of the
preparation site and to damage to the tooth substance. Ensure that
the water content is ≥ 50 ml/min.
CAUTION!  Never pull the patient's cheek back with the
handpiece! This would actuate the pushbutton, thus creating a risk of
burning the patient's oral mucosa.
CAUTION!  Potentially hazardous optical radiation may cause
harm to the eyes. Do not stare at the LED for longer periods of time
while in operation.
IMPORTANT: For application details and operating data of the
different preparation tools, please refer to the information provided by
the manufacturer.
IMPORTANT: Prevent the development of infections and
emphysema from spray air penetrating into open wounds.
5 Operation
29