7 Reprocessing
44
1.
Fit the spray adapter onto the nozzle of the spray can.
2.
Insert the instrument until it clicks into place and hold it.
3.
Spray into the instrument for 1-2 seconds.
IMPORTANT: Hold the spray can upright.
4.
Wipe away any spray that runs out with a clean, lint-free cloth.
5.
Repeat the process until the spray that runs out is clear.
6.
Purge the instrument with compressed air (3 bar) for 10
seconds.
7.2.3
Lubricate the push button chuck
Use Midwest Plus Spray on the push-button chuck to remove
deposits and ensure proper functioning of the clamping system.
Instructions for Use
66 07 290  D3673.201.01.08.02