Instructions for Use
Preventing the spread of infections and cross
Coolant spray and wound contact
66 07 290  D3673.201.01.08.02
2
Safety notices
CAUTION!  Risk of contamination!
contamination
Prevent the spread of infection and cross contamination between
patients, users and third parties.
Before initial start-up and after every treatment, the product and
accessories must be reprocessed, see "Reprocessing" section.
Take the appropriate hygiene measures, e.g., wear protective gloves.
CAUTION!  Risk of injury! Important: Prevent the development of
infections and emphysema from spray air penetrating into open
wounds.
Filtered water
CAUTION!  Risk of burns! Use filtered water only (< 50 µm).
Sufficient cooling
CAUTION!  Risk of burns! Insufficient cooling leads to overheating
of the preparation site and to damage to the tooth substance. Ensure
that the water content is ≥ 50 mL/min.