Instructions for Use
66 07 290 D3673.201.01.08.02
6
Postprocessing
6.1
After each treatment session
CAUTION! Risk of contamination! Perform reprocessing
immediately after treatment, after 1 hour at the latest.
ü
The preparation instrument is at a standstill.
ü
Wear suitable protective clothing.
1.
Purge the water and air channels on the treatment center for 30
seconds.
2.
Remove the preparation instrument with tweezers.
3.
Remove the turbine.
4.
Transport the turbine to the hygiene room in a suitable transport
container.
5.
CAUTION! Risk of contamination! Conduct hygienic
reprocessing [→ 39]. Observe the relevant national/local
regulations.
6 Postprocessing
37