Instructions for Use
66 07 290  D3673.201.01.08.02
7.3
Sterilization
CAUTION!  Risk of contamination! Make sure you observe the
following intervals.
Intervals
Prior to initial operation
Prior to each subsequent use
Wrapped sterilization
ü
The high-speed handpiece is cleaned  [→ 39].
ü
The high-speed handpiece is lubricated.  [→ 41]
ü
The instrument is packed in packaging suitable for sterilization
and storage (FDA cleared, paper/plastic composite packaging
according to ISO 11607).
Sterilize the instrument in the steam sterilizer with saturated
water vapor.
7 Reprocessing
49