Instructions for Use
66 07 290 D3673.201.01.08.02
7.3
Sterilization
CAUTION! Risk of contamination! Make sure you observe the
following intervals.
Intervals
●
Prior to initial operation
●
Prior to each subsequent use
Wrapped sterilization
ü
The high-speed handpiece is cleaned [→ 39].
ü
The high-speed handpiece is lubricated. [→ 41]
ü
The instrument is packed in packaging suitable for sterilization
and storage (FDA cleared, paper/plastic composite packaging
according to ISO 11607).
➢
Sterilize the instrument in the steam sterilizer with saturated
water vapor.
7 Reprocessing
49