Procedure
Tip: Record the time and result of the check for your own information.
ü
The expiry date of the chuck tester (A) has not elapsed.
1.
Insert the chuck tester into the FG clamping system.
2.
Tighten the chuck tester until the marking ring appears
(withdrawal force: 22 N).
3.
Check if the chuck tester slips out of the chuck before the
marking ring appears.
Ä
If no: The chuck is functional.
Ä
If yes:
CAUTION!  Risk of injury!  The chuck is defective
and the secure fit of the preparation tool cannot be
guaranteed.
Stop using the high-speed handpiece.
Have the high-speed handpiece's clamping system
replaced in a workshop authorized by Midwest.
8 Maintenance
53