Instructions for Use
66 07 290 D3673.201.01.08.02
After sterilization
1.
Remove the high-speed handpiece from the steam sterilizer.
CAUTION! The high-speed handpiece may still be hot. Do
not accelerate the cooling off process up by plunging the high-
speed handpiece in cold water. This will damage your high-
speed handpiece!
2.
Perform a visual check to see if the packaging is dry after
sterilization.
3.
Store all high-speed handpieces so that they are protected from
contamination.
4.
Sterilize again once the storage period has elapsed.
7 Reprocessing