13-01811 | Thermocouple Meter/Calibrator | Tegam | SN: 2454186 | Cal: 10/04/2023 | Due: 02/28/2024 | Range: 2400 °F | Report #: 31363-217443-3646
LP-01782 | Thermo-Hygrometer | Comark | SN: 06247790052 | Cal: 01/24/2024 | Due: 01/31/2025 | Range: 122 °F 95 %RH | Report #: 32568-205513-3646
Calibration Date:
Recommended Due Date:
Calibration Frequency:
Manufacturer:
Type:
Model Number:
Serial #:
Resolution:
Capacity:
Tolerance:
Thermocouple Type:
Channel
Bottom
Channel
Meter B
Data Logger 10 Channel 6 Point (PCC) CF-013-03
Page: 1 of 2
Report and Certificate of Calibration
33086-206391-4525-B
PFS TECO
1507 Matt Pass
Cottage Grove
Ethan Frederick
11785 SE Highway 212, Suite 305
Calibration Standards
February 26, 2024
February 26, 2025
12 Months
National Instruments
Data Logger
NI 9213
1B182FB
0.1 °F
2,500 °F
±
3.0 °F
K
LOGGING FUNCTION
Calibration
UUT
Standard
As Found
0.00
-0.10
500.00
499.90
1000.00
1000.00
1500.00
1500.10
2000.00
2000.10
2400.00
2400.00
0.00
-0.10
LOGGING FUNCTION
Calibration
UUT
Standard
As Found
0.00
-0.10
500.00
499.90
1000.00
1000.10
1500.00
1500.10
2000.00
2000.20
2400.00
2400.20
0.00
0.00
Copyright 2013 Cal-Cert. All rights reserved.
Customer PO#:
State: WI
Clackamas, OR 97015
Instrument Data
Reference:
Cal-Cert Procedure:
Indicating System:
Temperature:
Humidity:
Asset #:
Service Location:
As Found:
As Left:
Thermocouple
UUT As
UUT As
As Left
Left
Left
Error
Reading 1
Reading 2
-0.10
-0.10
0.10
499.90
499.90
-0.10
1000.00
1000.00
0.00
1500.10
1500.10
0.10
2000.10
2000.10
0.10
2400.00
2400.00
0.00
-0.10
-0.10
0.10
Thermocouple
UUT As
UUT As
As Left
Left
Left
Error
Reading 1
Reading 2
-0.10
-0.10
0.10
499.90
499.90
-0.10
1000.10
1000.10
0.10
1500.10
1500.10
0.10
2000.20
2000.20
0.20
2400.20
2400.20
0.20
0.00
0.00
0.00
1109
Zip: 53527
Navair 17-20ST-95
CP-013
Computer
66 °F
34% RH
215 Booth 1
Service Address
Pass
Pass
Expanded
Uncertainty±
0.346
Expanded
Uncertainty±
0.346
Revision 12
7/24/2017