INSTRUCTION MANUAL
APPENDIX C ELECTROMAGNETIC COMPATIBILITY
MD300K2 conforms to IEC60601-1-2:2014 Electromagnetic Compatibility (EMC) standard.
Essential performance is defined as SpO
of abnormal operation. Accuracies may be affected as a result of exposure to electromagnetic
disturbances that are outside of the environments listed in the intended use. If issues are
experienced, move the device away from the source of electromagnetic disturbances.
Table 1: Electromagnetic Emissions Limits and Compliance
Emissions Test
RF Emissions CISPR 11
Note: Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3) are
not applicable.
Table 2: Electromagnetic Immunity
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Rated power Frequency Magnetic Fields
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Conduced RF
IEC 61000-4-6
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Software Downloding
Please visit the websites below and find the service center to download the software for PC.
http://www.choicemmed.com
http://www.choicemmed.eu
http://www.choicemmed.in
http://www.choicemmedamerica.com
accuracy and pulse rate accuracy or an indication
2
Compliance
Group 1, Class B
Compliance
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
30 A/m
50Hz and 60 Hz
80 MHz – 2.7 GHz
±2 kV for power supply lines
±1 kV differential mode
3 V 0,15 MHz – 80 MHz
6 V in ISM bands between 0,15 MHz and 80 MHz
0 % UT; 0,5 cycle at 0°, 45°, 90°,135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30 cycles;
Single phase: at 0°
0 % UT; 250/300 cycle
- 27 -
MD300K2
10 V/m 80% AM 1kHz