MD300K2
PULSE OXIMETER
5 SETTING
5.1 System Setting
Pick and enter the [System Setting] interface from the main menu. And then press the
Navigation buttons to select different item to set.
Fig.5.1
Fig.5.2
Alarm Volume: you can adjust the value of alarm volume, there are 7 levels, and the default
level is 3.
Alarm Pause: there are two modes, 60s and 120s, and the default mode is 120s. The device
audibly alarm upon new alarm conditions.
Beep Tone: the level is from 0 to 7, and the default level is 3.
Backlight Setting: the level of brightness is from 1 to 7, and the default level is 3.
Language: English, French, German, Spanish, Italian, Japanese, Russian and Chinese.
Screen Sleep Mode: 1minute, 10 minutes, 30 minutes, screen always on, and the default is
1 minutes
Charging management: Charging activated, Charging stop.
ID setup: ① press OK button, ② press Navigation button to change the number, ③ press OK
button to confirm.
Factory Default: recover to factory reset.
Note:
1. Every time enter into the Alarm Volume and Alarm Pause, you should input the password,
refer to Alarm Setting.
2. The ID range is 1~127.