Installation and Operation Manual
X-VAP-VDM300-eng
Part Number: 541B210AAG
October, 2018
2-11 Product Safety Information
2-9
Brooks
Make sure you understand the following safety information before you install
and operate the VDM300.
Hazards
The VDM300 does not use hazardous chemicals or gases in the normal
operation or maintenance of the unit. Nitrogen is used during the drain cycle
and de-ionized water is used as the source for creating water vapor.
Safety Interlocks
The VDM300 uses 208-230 VAC to power the tank heaters. The tank heat-
ers have a primary protection of two fuses located in the fuse drawer of the
power entry module. The fuses are rated for 3 amps at up to 250 VAC. A
secondary form of protection for the tank heaters is a Thermal Cut-Off fuse
(TCO) in series with the heaters.
Outlet Fitting
The outlet fitting will become hot during normal operation. Take the neces-
sary precautions to avoid being burned such as wearing gloves. Refer to the
hazard label on the VDM300.