Télécharger Imprimer la page

Brooks Instrument VDM300 Manuel D'installation Et D'utilisation page 23

Publicité

Brooks
VDM300 Vapor Delivery Module
®
2-7 Diagnostic Port Connection
2-8 Return Shipment
2-9 Transit Precautions
2-10 Removal from Storage
The VDM300 can also be operated in a Service Mode through the micro USB
diagnostic port connection utilizing Brooks Expert Support Tool (BEST) and a
USB to micro USB cable. You should not use this port for normal operation.
The BEST GUI software software installed on a PC acts as the interface to
perform these operations. For more information on the use of the BEST, see
Section 4-5 Operation of the Brooks Expert Support Tool (BEST) for VDM300
later in this manual and/or the BEST manual available on the Brooks Instru-
ment website.
Prior to returning any device to the factory, visit the Brooks web site (www.
BrooksInstrument.com) for a Return Materials Authorization Number (RMA#),
or contact one of the locations provided on p. 2-1
Prior to returning the device, it must be drained & purged in accordance with
the following:
All flow devices returned to Brooks require completion of the Brooks Instru-
ment Decontamination Statement, along with a Safety Data Sheet (SDS)
for the fluid(s) used in the instrument. Failure to provide this information will
delay processing by Brooks personnel.
Copies of these forms can be downloaded from the Brooks website (www.
brooksinstrument.com/globalsupportcenters) or are available from any of the
Brooks Instrument locations provided on the back cover.
To safeguard against damage during transit, transport the device to the
installation site in the same container used for transportation from the factory,
if circumstances permit.
Upon removal of the device from storage, a visual inspection should be
conducted to verify its "as-received" condition. If the device has been subject
to storage conditions in excess of those recommended (refer to "2-3 Recom-
mended Storage Practice" on p. 16), it should be subjected to a pneumatic
pressure test in accordance with applicable vessel codes.
Installation and Operation Manual
X-VAP-VDM300-eng
Part Number: 541B210AAG
October, 2018