4.1.3. Attaching the Patient Tracker Advanced to the base
1.
Preassemble the Patient Tracker Advanced and attach it to the Universal Base
Skull. Refer to the Patient Tracker Advanced Instructions for Use.
WARNING
Fix the navigation spheres to the Patient Tracker Advanced before attaching it to the
–
Universal Base Skull.
Hold the Universal Base Skull firmly while attaching the Patient Tracker Advanced to
–
the base so that the base neither twists nor tilts on the patient.
4.2. Instructions for application
WARNING
Do not exert force when using the patient tracker for instrument validation.
14/54 | Using the product