Caution
When you dispose of the unit or accessories, you
must obey the laws in the relative area or country
and the regulations in the relative hospital
regarding environmental pollution.
Warning on power connections
Use a proper power cord for your local power
supply.
1. Use the approved Power Cord (3-core mains
lead) / Appliance Connector / Plug with
earthing-contacts that conforms to the safety
regulations of each country if applicable.
2. Use the Power Cord (3-core mains lead) /
Appliance Connector / Plug conforming to the
proper ratings (Voltage, Ampere).
If you have questions on the use of the above
Power Cord / Appliance Connector / Plug, please
consult a qualified service personnel.
Warning on power connections for
medical use
Customers in the U.S.A. and Canada should use
the following type of power cord.
Customers in other countries or regions should
use the power cord prescribed by their country or
region.
Plug type
Cord type
Minimum rating for
plug and appliance
couplers
Safety approval
* Note: Grounding reliability can only be achieved
when the equipment is connected to an
equivalent receptacle marked “Hospital Only” or
“Hospital Grade”.
Warning
The apparatus shall not be exposed to dripping or
splashing. No objects filled with liquids, such as
vases, shall be placed on the apparatus.
U.S.A. and Canada
*
HOSPITAL GRADE
Min. Type SJT
Min. 18 AWG
10 A / 125 V
UL Listed and CSA
Warning
Make sure the surface is wide enough so that this
apparatus’s width and depth don’t exceed the
surface’s edges.
If not, this apparatus may lean or fall over and
cause an injury.
Warning
To prevent injury, if mounting the unit using a
mounting arm, wall fixture, or other mounting
device prepared by the customer, mount the unit
securely as described in the instruction manual
provided with the mounting device.
Check beforehand that the mounting device used
has sufficient strength to support the added
weight of the unit.
Check yearly that the mounting device is securely
attached.
Caution
When installing, ensure the following space
around the periphery of the unit, taking
ventilation and servicing into consideration.
– Rear side: 10 cm (4 in.) or more
– Left/Right sides: 10 cm (4 in.) or more
– Bottom side: 5 cm (2 in.) or more
– Top side: 30 cm (11
7
/
8
Consult with Sony qualified personnel for the
following types of installation location.
– Wall mount
– Mounting arm
Caution
SERIAL REMOTE connector (RJ-45)
For safety, do not connect the connector to
peripheral device wiring that might have
excessive voltage.
Follow the instructions for use for this port.
Caution
SERIAL REMOTE connector (RJ-45)
Do not touch this connector and a patient at the
same time.
Doing so may result in a generation of voltage that
can be harmful to patients if the unit is
malfunctioning.
Always disconnect the power cord before
connecting and disconnecting connectors.
in.) or more
9
GB