Warning
Using this unit for medical purposes
The connectors on this unit are not isolated.
Do not connect any device other than one which
conforms to IEC 60601-1 standards.
When an information technology device or AV
device that uses an alternating current is
connected, current leakage may result in an
electric shock to the patient or operator.
If use of such a device is unavoidable, isolate its
power supply by connecting an isolation
transformer, or by connecting an isolator between
the connecting cables.
After implementing these measures, confirm that
the reduced risk now conforms to IEC 60601-1
standards.
Caution
Do not use the device in a MR (Magnetic
Resonance) environment.
It may cause a malfunction, fire, and unwanted
movement.
Caution
This unit with the cable covers attached conforms
to the waterproofed standard. (See page 16)
Be sure not to use the unit with the cable covers
removed since waterproofed performance is not
guaranteed.
Caution
This equipment is not suitable for use in locations
where children are likely to be present.
Caution
• This unit is heavy. Make sure to unpack and
move the unit with two or more people.
• Firmly grip the bottom of this unit as shown
below.
10
GB
For details on operation, refer to the manuals
stored on the supplied CD-ROM.
Using the CD-ROM manuals
The manuals can be viewed on a computer with
Adobe Reader installed.
You can download Adobe Reader for free from
the Adobe website.
1. Open the “index.html” file on the CD-ROM.
2. Select the language of the manual you want
to view.
Note
If you damage or lose the CD-ROM, you can
purchase a new one from your dealer or Sony
service representative.
Reduction in the Use of Hazardous Substances
in Electrical & Electronic Equipment (Applicable
in Republic of India)
This product and its components, consumables,
parts or spares comply with the hazardous
substances restriction of India’s E-Waste
(Management) Rules. The maximum allowable
concentrations of the restricted substances are
0.1% by weight in homogenous materials for Lead,
Mercury, Hexavalent Chromium, Polybrominated
Biphenyls (PBB) and Polybrominated Diphenyl
Ethers (PBDE), and 0.01% by weight in
homogenous materials for Cadmium, except for
the exemptions specified in Schedule II of the
aforesaid Rules.

Publicité

loading