The EyeSeeCam in not intended for operation in oxygen rich environment, flammable
anesthetics, flammable agents or similar
The goggle should not be worn by patients with strong defective vision and abnormal rare blink.
•
Please consult a specialist in such circumstances before using the mask on these types of
patients.
The system must be switched off before cleaning
•
•
When used for longer time in environments temperatures above 30
the camera unit can exceed 43
the equipment for a while.
•
The EyeSeeCam system must be serviced at least once a year. The service must include a
safety test. No parts can be serviced while in use with the patient.
Do not use the equipment if the enclosure is not intact, but have the equipment serviced.
•
Only personnel with proper training (skilled personnel) should operate the system.
•
Proper use of this device depends on careful reading of this manual and all additional instructions
•
and labels
•
Let the system obtain room temperature before turning it on. Extra time for obtaining room
temperature may be needed to avoid condensation if the system has been moved from a very
cold location to a warmer one.
The patients or test persons should not wear any optical utilities such as glasses or contact
•
lenses.
The patients or test persons should not wear makeup around the eyes
•
EyeSeeCam vHIT - Instructions for Use - EN
°
°
C (max 44
C). If this feels uncomfortable for the patient turn off
°
C the surface temperature of
Page
3